Product Deatails | |
---|---|
Flow | 300 LPH (MAX) |
Gas services | Air ,N2, H2 ,He, Ace |
Inlet connection | 1/4" BSP male to suit pigtail |
Inlet Pressure Gauge | 2 nos. of 0 - 210 Kg Gauge |
Outlet Connection | 1/4"OD ferrule fitting with ball valve |
Outlet Pressure Gauge | 1 no. of 0 - 14 Kg Gauge. |
Usage | Laboratory |
Setting Procedure:
- Open outlet valve to use as per requirement
- Set the outlet pressure as per requirement With outlet press Regulator
- Set the pressure through cylinder 2 regulator ( e..g. 6 Kg.)
- Open the cylinder 2 and apply the pressure of cylinder 2 ( Line 2 ) to the manifold.
- Set the pressure through cylinder 1 regulator ( e.g. 7 kg)
- Open the cylinder 1 and apply the pressure of Cylinder 1 ( Line 1) to the manifold
- Connect both cylinders with flexible pigtails